SEA1200VX High Sensitivity
Element Monitor

XRF Analyzer suitable for RoHS&ELV analysis

Name
High Senditivity Element Monitor
Type
SEA1200VX
Overview
Meeting both demands for screening inspection for compliance with RoHS/ELV and advanced analysis. Equipped with the unique "Vortex" XRF detector featuring high resolution, high sensitivity and high count rate, SEA1200VX has remarkable sensitivity. Precision control software greatly shortens measurement time.
Icons
SEA6000VX

Phenominal High Speed Analysis (High Sensitivity)

Measuring time can be greatly shortened to an average one tenth the time of conventional method by combining the high count rate detector and precision control software.

LN2 Less High Count Rate Detector

Newly equipped world's best class high count rate (150,000cps) detector "Vortex" that does not require LN2 greatly surpasses conventional detectors in sensitivity and resolution.

Newly Developed Uni Filter (Optional)

Five RoHS restricted elements can be measured at one time using a single composite filter, achieving even shorter measuring times than the already rapid measuring time of the SEA1200VX.

Wide Range of Application

Measurement of Na to U element is possible by conducting measurement in vacuum atmosphere. (Light Element Option) Also supports highly sensitive analysis of Cl.

Micro Spot Analysis

Micro spot qualitative and quantitative analysis with a 1mm X-ray beam anywhere on a printed circuit board.

Sample Changer (Optional)

Enables continuous measurement of up to 12 samples that require different measurement conditions. Large samples of up to 430 x 320mm in size can also be measured sinply by being placed in the chamber.

Elements
Atomic nos. 11 (Na) to 92 (U)
*Atomic nos. when using Vacuum Option
Sample State
Solid / Powder / Liquid
X-ray source
Small Air-cooled X-ray tube (Rh target)
Voltage: 15kV, 30kV, 40kV, 50kV
Current: 1mA
X-ray direction
Bottom-up Irradiation
Detector
Vortex Si semiconductor detector (No liquid nitrogen required)
Analysis area (beam size)
Square 0.2mm, 0.5mm, 1.2mm, 3mm 
Electric switching
Electric switching
Sample Observation
Color CCD camera
Chamber
430 (W) x 320 (D) X 200 (H) mm
239 (W) x 280 (D) X 66 (H) mm (When using vacuum)
Filter
5 filters automatic switching
Controller
Laptop or Desktop
Qualitative functions
Spectrum measurement, Auto-ID, Comparison display
Quantitative functions
KLM marker display, Subtraction display, Bulk (FP) CAL, Calibration
Data Process
Microsoft® Excel, Microsoft® Word
Power requirements (Electric box included)
AC100V to 120V, 200-240V ±10%, 5A.
• HSEASY (Precision control software) 
• Spectrum Matching Software (Material Discrimination) 
• Film Analysis (CAL) Software
• Sample Changer (turret)
• 3mm Colimeter
• Vacuum Pump (Light element analysis)
• Standard Reference Samples 
• UNI Filter
*Microsoft is a registered trademark of Microsoft Corporation in the United States and other countries.
Contact Us
Calibration
Coating Thickness
Standards
Cold Chain
Management
X-Ray
Analysis & Safety
XRF Machines
& Standards